BIO Safety Cabinetx
Specifications
- Bio Safety Type: 70% Re-Circulation, 30% Exhaust Through HEPA
- Cleanliness: ISO CLASS 5 or CLASS 100 or M 3.5 Area
- Velocity at Grill: 90 20 FPM
- Construction: Models available in 18 gauge (1.2 mm) stainless steel 304 ( Internal & External ).
- HEPA Filter: HEPA Filters down to 0.3mic. particulate at an efficiency of 99.997% duly tested on
- DOP aerosol method
- SIZE: 1no – 915 X 610 X 69 at Supply/Equivalent
- SIZE: 1no – 610 X 305 X 68 at Exhaust/Equivalent
- Noise level: 65 DB measured from filter face.
- SEMI HEPA Filter: HEPA Filters down to 0.3mic. Particulate at an efficiency of 50% duly tested on DOP aerosol method
- SIZE: 1no – 610 X 610 X 68 at Section/Equivalent
- Work Top: Perforated Work SS-304 Top with Train Dray
- Front Sash: Sliding Door with Counter weight
- Accessories: GAS / AIR cock – 01 No. 2.DOP Port
- Electrical: ON/OFF Panel – (BL/UV/FL)
- Diffused Fluorescent Light – 36W to maintain 800lux – 1No. (Make: Philips)
- U V light with Hour meter – 30W – 1No – (Make: Philips)
- Auxiliary Socket: 15 Amps – 1 No.
- Pressure Gauge: 1No. Magnehelic Gauge ( Make: Omicron )
- Power Supply: 230 V, 50Hz , Single Phase